Agilent Varian Turbo-V 70D 4.5" CF - REBUILT
Weight | 4 lbs |
---|---|
Pumping Speed | 60 l/s |
Ultimate Pressure | <1 x 10-9 mbar |
Agilent Varian Turbo-V 70D Turbomolecular Vacuum Pump - REBUILT
Each turbo pump must match its controller in reference to the same class. For example the TV 550 pump must match a V550 (stand alone, brick or PCB) controller. We also offer repair service.
The Turbo-V 70D pump consists of a high frequency motor driving a turbine fitted with 7 bladed stages and 4 Macrotorr stages. Please notice our special pricing for the ISO-63 Turbo-V 70D turbo pump with free US ground shipping. Other options are available with pricing on our Config page.
Questions? Please call our toll free number at 866-332-0500 or send an email to our sales team!
Specifications | Turbo-V 70D | |
---|---|---|
Pumping speed (l/s) | w/ 4.5" CF or ISO-63: N2: 60 l/s He: 30 l/s H2: 20 l/s |
w/ 2.75" CF or KF 40 NW: N2: 40 l/s He: 27 l/s H2: 20 l/s |
Compression ratio | N2: 5 x 107 He: 7 x 103 H2: 1.5 x 103 |
|
Base pressure* | w/ recommended mechanical forepump: 1 x 10-9 mbar (8 x 10-10 Torr) w/ recommended diaphragm forepump: 1 x 10-9 mbar (8 x 10-7 Torr) |
|
Inlet flange | 4.5" O.D. CF ISO-63 2.75" O.D. CF KF-40 NW |
|
Foreline flange | KF-16 NW | |
Rotational speed | 75,000 rpm | |
Start-up time | < 60 seconds | |
Recommended forepump | mechanical: Varian SD-40 diaphragm: Varian MD-6 2 m3/h (1.2 cfm) |
|
Operating position | any | |
Cooling requirements | Natural air convection Forced air or water optional |
|
Operating ambient temperature | + 5°C to + 35°C | |
Coolant water | flow: 10 l/h (0.05 GPM) temperature: + 10°C to + 30°C pressure: 2 to 4 bar (30 to 60 Psi) |
|
Bakeout temperature | 120°C at inlet flange max. (CF flange) 80°C at inlet flange max. (ISO flange) |
|
Vibration level (displacement) | < 0.01 μm at inlet flange | |
Noise level | 45 dB (A) at 1 meter | |
Input | 54 Vac, three-phase, 1,250 Hz | |
Lubricant | Permanent lubrication | |
Installation category | II | |
Pollution degree | 2 | |
Storage temperature | - 20°C to + 70°C | |
Weight kg (lbs) | ISO: 2.3 (6) CF: 1.6 (4) |