Leybold TURBOVAC 361 DN 160 CF - REBUILT

Call for Price

In stock
SKU
RB_TMP361-85673
Weight 26.00 lbs
Pumping Speed 345 l/s
Ultimate Pressure < 1 x 10-10 mbar

Leybold Vacuum TURBOVAC 361 Turbomolecular Pump - REBUILT

Most configurations are being sold discounted while supplies last.

Sold with one year warranty. The 361C corrosive version is also available; please call for a quote and lead time. We offer repair service and are capable of rebuilding, balancing and testing this pump. View our inventory of Leybold Vacuum turbo pumps.

Typical Applications

  • Leak detectors
  • Mass spectrometers
  • Data storage
  • Optical coating
  • R & D: UHV systems, Particle accelerators
  • Load locks and transfer chambers

Technical Features

  • Small footprint
  • Installation in any orientation
  • Oil-free pump for generating clean high and ultrahigh vacuum conditions
  • Bearing temperature measurement through the TURBO.DRIVE TD 20 classic

Advantages to the User

  • Space-saving
  • Easy to integrate into complex vacuum systems
  • Low operating costs
  • Highly reliable operation also in processes loaded with particles

Questions? Please call our toll free number at 866-332-0500 or send an email to our sales team!

Inlet flange DN 100 ISO-K • 100 CF DN 160 ISO-K • 160 CF
Pumping speed
N2 l/s 345 400
Ar l/s 350 -
He l/s 340 380
H2 l/s 340 370
Max. gas throughput
N2 mbar x l/s 3.0
Ar mbar x l/s 2.5
Compression ratio
N2 1 x 109
He 6 x 104
H2 3 x 103
Ultimate pressure mbar (Torr) < 1 x 10-10 (< 0.75 x 10-10)
Ultimate partial pressure mbar < 10-10
Max. continuous inlet pressure mbar (Torr) 5 x 10-2 (4 x 10-2)
Max. foreline pressure for N2 mbar (Torr) 5 x 10-1 (4 x 10-1)
Foreline connection KF25, ISO 63
Mounting position Any
Bearing type Grease/ceramic ball bearing
Corrosive version (C) Yes
Cooling Standard: Water • Available: Air
Recommended forevacuum pump Standard: TRIVAC D 16 B • Corrosive: D 25 B
Run-up time at 95% speed 2 min
Purge / vent port DN 10 KF
Cooling water connection (hose nozzles) 10 mm (0.39 in)
Weight 12 kg (26 lbs)
Max. power consumption / Ultimate pressure 680 / 480 VA
Frequency converter TD 20, NT 20, NT 150/360

Specifications

Need modifications to this item? PTB Sales' expert team of engineers can configure products to your specific needs.

Leybold Vacuum TURBOVAC 361 Turbomolecular Pump - REBUILT

85670 / 856 70  TURBOVAC 361  DN 100 ISO-K
85671 / 856 71  TURBOVAC 361  DN 100 CF
85672 / 856 72  TURBOVAC 361  DN 160 ISO-K
85673 / 856 73  TURBOVAC 361  DN 160 CF
85674 / 856 74 TURBOVAC 361 DN 100 ISO-K SCIEX 85450 / 854 50 TMP-360 DN 100 ISO-K 85451 / 854 51 TMP-360 100 85456 / 854 56 TMP-360 DN 160 CF 89577 / 895 77 TMP-360 4" ASA 85515-1 / 855 15-1 TMP-360C DN 100 ISO-K